Abstract
Two ambient-temperature, dry plasma transport systems, ViveST tubes and RNASound RNA sampling cards, and two extraction methods were compared to frozen plasma for HIV-1 RNA recovery. Significant RNA loss occurred: ViveSTMiniMag> ViveSTQIAamp>RNASoundQIAamp. RNA loss and low specimen volumes may affect the sensitivity of genotyping specimens with HIV-1 RNA of<4.70 log10 copies/ml.
Cite
CITATION STYLE
Levine, M., Beck, I., Styrchak, S., Pepper, G., & Frenkel, L. (2016). Comparison of matrix-based and filter paper-based systems for transport of plasma for HIV-1 RNA quantification and amplicon preparation for genotyping. Journal of Clinical Microbiology, 54(7), 1899–1901. https://doi.org/10.1128/JCM.00541-16