Abstract
This article explains a step-wise protocol to develop an electrochemical sensor to quantify serum levels of C-telopeptide (CTX) crosslinks also known as crosslaps in a matter of minutes and with high level of accuracy. The technique needs only one-step (incubation) and can thus be used for point of care screening. Due to the excellent electrical properties of the as-prepared immunosensor, CTX levels were successfully measured from 1 to 1000 pg/mL. This is while the normal reference of the marker is 50–450 pg/mL, suggesting that the sensor can acceptably detect CTX. The results also showed a good correlation with ECLIA in measuring serum levels of CTX.
Author supplied keywords
Cite
CITATION STYLE
Khashayar, P., Amoabediny, G., Larijani, B., Hosseini, M., Verplancke, R., De Keersmaecker, M., … Vanfleteren, J. (2017). A highly sensitive electrochemical biosensor based on AuNP-modified gold electrodes for selective determination of serum levels of crosslaps. 3 Biotech, 7(5). https://doi.org/10.1007/s13205-017-0919-y