Safety and performance of a suprachoroidal sensor for telemetric measurement of intraocular pressure in the EYEMATE-SC trial

8Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

Abstract

Aim To investigate the safety and performance of a telemetric suprachoroidal intraocular pressure (IOP) sensor (EYEMATE-SC) and the accuracy of its IOP measurements in open angle glaucoma (OAG) patients undergoing simultaneous non-penetrating glaucoma surgery (NPGS). Methods Prospective, multicentre, open-label, single-arm, interventional clinical trial. Twenty-four eyes of 24 patients with OAG regularly scheduled for NPGS (canaloplasty or deep sclerectomy) were simultaneously implanted with an EYEMATE-SC sensor. Six-month follow-up on the sensor’s safety and performance as well as on the level of agreement between the EYEMATE-SC measurements and IOP measurements with Goldmann applanation tonometry (GAT). Results The eyes underwent canaloplasty (n=15) or deep sclerectomy (n=9) and achieved successful implantation of the sensor. No device migration, dislocation or serious device-related complications occurred. A total of 367 comparisons were included in the IOP agreement analysis. The overall mean difference between GAT and EYEMATE-SC measurements was 1.31 mm Hg (lower limit of agreement (LoA) 7.55 mm Hg; upper LoA –4.92 mm Hg). The maximum difference of 2.5 mm Hg ±3.96 (LoA 0.30–2.29) was reached on day 10 and continuously improved to an agreement of –0.15 mm Hg ±2.28 (LoA –1.24 to 0.89) after 6 months. Accordingly, the percentage of eyes within an IOP difference of ±5 mm Hg improved from 78% (day 3) to 100% (6 months). Conclusions After 6 months, the EYEMATE-SC sensor was safe and well tolerated, and allowed continual IOP monitoring.

Cite

CITATION STYLE

APA

Szurman, P., Mansouri, K., Dick, H. B., Mermoud, A., Hoffmann, E. M., Mackert, M., … Seuthe, A. M. (2023). Safety and performance of a suprachoroidal sensor for telemetric measurement of intraocular pressure in the EYEMATE-SC trial. British Journal of Ophthalmology, 107(4), 518–524. https://doi.org/10.1136/bjophthalmol-2021-320023

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free