An oropharyngeal pH monitoring device to evaluate patients with chronic laryngitis

41Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Diagnostics for gastro-esophageal reflux disease (GERD) are suboptimal because of limited sensitivity. We performed in vitro and in vivo studies to systematically assess the performance characteristics of an oropharyngeal pH probe. Methods: In vitro studies compared the oropharyngeal probe with a standard pH catheter in liquid and aerosolized solutions, pH 1-7. The accuracy of measurements, deviation from target pH, and time to equilibrium pH were determined. Simultaneous distal esophageal pH measurements were obtained in 11 patients with GERD. Oropharyngeal and distal esophageal reflux parameters were measured for controls (n = 20), patients with GERD (n = 17), and patients with chronic laryngitis (n = 10). Key Results: In the liquid phase, at pH 4-5, the oropharyngeal probe had less deviation from the target value than the standard catheter; deviation in the vapor phase was similar (0.4 pH units). Median (interquartile) time to reach equilibrium pH was significantly (P < 0.001) faster with the oropharyngeal than the standard probe. In comparing simultaneous distal esophageal pH characteristics, 96% of recordings with the new and standard probes were in agreement to within ±1.0 pH unit; 71% of recordings were in agreement within ±0.5 pH units. Patients with chronic laryngitis had significantly higher levels of oropharyngeal acid exposure at pH <4, <5, and <6, in the upright position than patients with GERD or controls (P < .001). Conclusions & Inferences: Oropharyngeal pH monitoring appears to be more sensitive than traditional pH monitoring in evaluation of patients with extraesophageal reflux. It is a promising tool in evaluation of this difficult group of patients. © 2013 Blackwell Publishing Ltd.

Cite

CITATION STYLE

APA

Yuksel, E. S., Slaughter, J. C., Mukhtar, N., Ochieng, M., Sun, G., Goutte, M., … Vaezi, M. F. (2013). An oropharyngeal pH monitoring device to evaluate patients with chronic laryngitis. Neurogastroenterology and Motility, 25(5). https://doi.org/10.1111/nmo.12109

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free