Abstract
Purpose To investigate the accuracy and noise performance of fat quantification with multiple gradient-echo images acquired using bipolar read-out gradients and compare them with those of the well-established unipolar technique. Theory The bipolar read-out technique induces phase and amplitude errors caused by gradient delays, eddy currents, and frequency-dependent coil sensitivity. In this study, these errors were corrected for jointly with the fat/water separation by modeling the impact of these effects on the signal. This approach did not require acquisition of reference data or modification of the pulse sequence. Methods Simulations and a phantom experiment were used to investigate the accuracy and noise performance of the technique and compare them with those of a well-established technique using unipolar read-out gradients. Also, the in vivo feasibility was demonstrated for abdominal applications. Results The phantom experiment demonstrated similar accuracy of the bipolar and unipolar fat quantification techniques. In addition, the noise performance was shown not to be affected by the added estimations of the phase and amplitude errors for most inter-echo times. Conclusion The bipolar technique was found to provide accurate fat quantification with noise performance similar to the unipolar technique given an appropriate choice of inter-echo time. © 2013 Wiley Periodicals, Inc.
Author supplied keywords
Cite
CITATION STYLE
Peterson, P., & Månsson, S. (2014). Fat quantification using multiecho sequences with bipolar gradients: Investigation of accuracy and noise performance. Magnetic Resonance in Medicine, 71(1), 219–229. https://doi.org/10.1002/mrm.24657
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.