Abstract

The patella clunk syndrome is not an uncommon complication, particularly after posterior-stabilized (PS) total knee arthroplasty. The present study was a consecutive unselected series comparing the incidence of the patella clunk syndrome amongst two comparable groups of elderly patients with knee arthritis implanted with the Insall-Burstein (IB II) total knee system and the newer NexGen Legacy total knee prosthesis. While 7.5% of patients in the PS IB II group developed patella clunk syndrome, none occurred in the NexGen group (P=0.05). We believe the difference is due mainly to a change in the design of the prosthesis, especially the more posterior intercondylar box and femoral cam of the NexGen femoral component.

Cite

CITATION STYLE

APA

Ip, D., Wu, W. C., & Tsang, W. L. (2002). Comparison of two total knee prostheses on the incidence of patella clunk syndrome. International Orthopaedics, 26(1), 48–51. https://doi.org/10.1007/s00264-001-0316-2