Estimation of adult patient doses for chest X-Ray examinations and comparison with diagnostic reference levels (DRLS)

4Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The aim of this study was to evaluate the radiation doses to patient during chest (posterior anterior/and lateral) examinations. The study was performed in three public hospitals of I˙stanbul province with a total of 300 adult patients. Entrance surface dose (ESD) measurements were conducted on computed radiography, digital radiography and screen film system. ESD was estimated by using International Atomic Energy Agency (IAEA) model and Davies model which are the common indirect models. Results were compared with diagnostic reference levels from the European Commission, IAEA and National Radiological Protection Board. Although the results are compatible with the international diagnostic reference levels, they present variations between the hospitals. Dose variations for the same type of X-ray examination support the idea that further optimization is possible.

Cite

CITATION STYLE

APA

Baş Mor, H., Altinsoy, N., & Söyler, I. (2018). Estimation of adult patient doses for chest X-Ray examinations and comparison with diagnostic reference levels (DRLS). Radiation Protection Dosimetry, 182(3), 377–385. https://doi.org/10.1093/RPD/NCY076