Successful use of continuous flow ventricular assist device in a patient with mechanical mitral and aortic valve prosthesis without replacement or exclusion of valves

12Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.

Abstract

In patients with left-sided mechanical aortic prostheses, it is recommended that the mechanical valve be replaced with a bioprosthesis, or excluded, at implantation of left ventricular assist device (LVAD). As changes in flow across the valve leads to potential thromboembolic complications, mechanical valves within the native heart are a relative contraindication to LVAD therapy. We here describe a patient who had long-standing valvular cardiomyopathy with mitral Starr-Edwards mechanical valve (Edwards Lifesciences, CA, USA) and aortic bileaflet tilting disc (St Jude Medical, St Paul, MN, USA) where LVAD was placed without explantation of the mechanical heart valves. The patient was bridged successfully to transplantation without thromboembolic events. © 2010 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.

Cite

CITATION STYLE

APA

Krishan, K., Pinney, S., & Anyanwu, A. C. (2010). Successful use of continuous flow ventricular assist device in a patient with mechanical mitral and aortic valve prosthesis without replacement or exclusion of valves. Interactive Cardiovascular and Thoracic Surgery, 10(2), 325–327. https://doi.org/10.1510/icvts.2009.221036