Amperometric biosensor for oxalate determination in urine using sequential injection analysis

14Citations
Citations of this article
26Readers
Mendeley users who have this article in their library.

Abstract

An amperometric flow biosensor for oxalate determination in urine samples after enzymatic reaction with oxalate oxidase immobilized on a modified magnetic solid is described. The solid was magnetically retained on the electrode surface of an electrode modified with Fe (III)-tris-(2-thiopyridone) borate placed into a sequential injection system preceding the amperometric detector. The variables involved in the system such as flow rate, aspired volumes (modified magnetic suspension and sample) and reaction coil length were evaluated using a Taguchi parameter design. Under optimal conditions, the calibration curve of oxalate was linear between 3.0-50.0 mg·L -1, with a limit of detection of 1.0 mg·L -1. The repeatability for a 30.0 mg·L -1 oxalate solution was 0.7%. The method was validated by comparing the obtained results to those provided by the spectrophotometric method; no significant differences were observed.

Cite

CITATION STYLE

APA

Rodriguez, J. A., Hernandez, P., Salazar, V., Castrillejo, Y., & Barrado, E. (2012). Amperometric biosensor for oxalate determination in urine using sequential injection analysis. Molecules, 17(8), 8859–8871. https://doi.org/10.3390/molecules17088859