Abstract
The purpose of this study was to assess and compare the reproducibility of trabecular bone structure measurements of the distal radius at 1.5 and 3.0 T magnetic resonance imaging (MRI). Root mean square reproducibility errors ranged from 0.69% to 4.94% at 1.5 T MRI and from 0.38% to 5.80% at 3.0 T MRI. Thus, reproducibility errors of trabecular bone structure measurements are overall in an acceptable range and similar at 1.5 and 3.0 T MRI. Copyright © 2012 by Lippincott Williams & Wilkins.
Author supplied keywords
Cite
CITATION STYLE
Baum, T., Dütsch, Y., Müller, D., Monetti, R., Sidorenko, I., Räth, C., … Bauer, J. S. (2012). Reproducibility of trabecular bone structure measurements of the distal radius at 1.5 and 3.0 T magnetic resonance imaging. Journal of Computer Assisted Tomography, 36(5), 623–626. https://doi.org/10.1097/RCT.0b013e31825f9aa3