An investigation of the response of the radiochromic dosimeter PRESAGE™ to irradiation by 62 MeV protons

8Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Measurements of the 62 MeV proton beam at the Clatterbridge Centre for Oncology using the radiosensitive plastic PRESAGE™ have previously shown a dependence of the dosimeter sensitivity (dose-response slope) on the linear energy transfer (LET) of the ionising particles. This work focuses on a possible explanation in terms of track structure theory (TST). Experimental measurements of highly irradiated PRESAGE™ samples established the D37 parameter of the theory to be of the order of 1000 Gy. Initial attempts at applying the theory showed good agreement of the theoretical and experimental values of relative effectiveness, but more work is needed to verify the model and understand its different parameters. © 2010 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Nowais, S. A., Kacperek, A., Brunt, J. N. H., Adamovics, J., Nisbet, A., & Doran, S. J. (2010). An investigation of the response of the radiochromic dosimeter PRESAGETM to irradiation by 62 MeV protons. In Journal of Physics: Conference Series (Vol. 250, pp. 155–159). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/250/1/012034

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free