How to detect and correct myocardial creep in myocardial perfusion imaging using Rubidium-82 PET?

11Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Reliability of myocardial blood flow (MBF) quantification in myocardial perfusion imaging (MPI) using PET can majorly be affected by the occurrence of myocardial creep when using pharmacologically induced stress. In this paper, we provide instructions on how to detect and correct for myocardial creep. For example, in each time frame of the PET images the myocardium contour and the observed activity have to be compared to check for misalignments. In addition, we provide an overview of the functionality of commonly used software packages to perform this quality control step as not all software packages currently provide this functionality. Furthermore, important clinical considerations to obtain accurate MBF measurements are given.

Cite

CITATION STYLE

APA

Koenders, S. S., van Dijk, J. D., Jager, P. L., Ottervanger, J. P., Slump, C. H., & van Dalen, J. A. (2019). How to detect and correct myocardial creep in myocardial perfusion imaging using Rubidium-82 PET? Journal of Nuclear Cardiology, 26(3), 729–734. https://doi.org/10.1007/s12350-019-01650-x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free