Confiabilidad de un monitor automatizado para la medición de la presión arterial

  • Cardona-Müller D
  • Grover-Páez F
  • Guzmán-Saldivar V
  • et al.
N/ACitations
Citations of this article
21Readers
Mendeley users who have this article in their library.

Abstract

Reliability of an automatic monitor for blood pressure measurement Background A correct blood pressure (BP) measurement is essential for the diagnosis and control of high BP. Aim: To evaluate the agreement and repeata-bility of BP measurements with the OMRON HEM-7320-LA device compared to a mercury sphygmomanometer. Material and Methods: A cross-sectional study comparing BP measurements made by two randomly selected trained nurses and an automatic oscillometric device. The mercurial sphygmomano-meter was connected to the automated device via a "T" type connector and a dual-head stethoscope was used, allowing simultaneous measurements. The results were analyzed with one-factor analysis of variance, Bland-Altman's test, repeatability coefficient (RC), and intra-class correlation coefficient (ICC). Results: Forty-nine participants aged 56 ± 19 years were included. Nineteen had hypertension (38%). We did not observe a significant difference in either systolic (SBP) or diastolic blood pressure (DBP) pressure measurements between the observers and the device. The mean difference was-0.09 mmHg (95% confidence intervals (CI)-0.9 to 0.7) for SBP and-0.9 mmHg (95% CI-1.7 to-0.13) for DBP. The RC for SBP (6.2, 5.2 and 5.8 mmHg) and DBP (4.7, 4.2 y 5.2 mmHg) was similar between the observers and the device. The ICC for SBP was 0.990 (95% CI 0.983 to 0.995, p < 0.01) and 0.986 (95% CI 0.977 to 0.991, p < 0.01) for DBP. Conclusions: There was a high level of agreement and similar measurement repeatability in the measurements performed by the automatic device and the mercurial sphygmomanometer. No differences in BP measurements were observed. (Rev Med Chile 2018; 146: 190-195)

Cite

CITATION STYLE

APA

Cardona-Müller, D., Grover-Páez, F., Guzmán-Saldivar, V., Alanis-Sánchez, G. A., Murguía-Soto, C., Totsuka-Sutto, S. E., … Ramos-Becerra, C. G. (2018). Confiabilidad de un monitor automatizado para la medición de la presión arterial. Revista Médica de Chile, 146(2), 190–195. https://doi.org/10.4067/s0034-98872018000200190

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free