Abstract
This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (-0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (-1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration. Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term. ©2010 British Editorial Society of Bone and Joint Surgery.
Cite
CITATION STYLE
McCalden, R. W., Charron, K. D., Yuan, X., Bourne, R. B., Naudie, D. D., & MacDonald, S. J. (2010). Randomised controlled trial comparing early migration of two collarless polished cemented stems using radiostereometric analysis. Journal of Bone and Joint Surgery - Series B, 92(7), 935–940. https://doi.org/10.1302/0301-620X.92B7.24462
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.