Abstract

Preanalytic laboratory testing errors are often difficult to identify. We demonstrate how laboratories can integrate statistical models with clinical judgment to develop protocols for preanalytic error detection. Specifically, we developed a protocol to identify spuriously elevated glucose values resulting from improper "line draws" or related phlebotomy errors. Using a decision tree-generating algorithm and an annotated set of training data, we generated decision trees to classify critically elevated glucose results as "real" or "spurious" based on available laboratory parameters. Decision trees revealed that a 30 -day patient-specific average glucose concentration lower than 186.3 mg/dL (10.3 mmol/L), a current glucose concentration higher than 663 mg/dL (37 mmol/L), and an anion gap lower than 16.5 mEq/L (16.5 mmol/L) suggested a spurious result. We then used the results from the decision tree analysis to inform the implementation of a clinical protocol that significantly improved the laboratory's identification of spurious results. Similar approaches may be useful in developing protocols to identify other errors or to assist in clinical interpretation of results. Copyright© by the American Society for Clinical Pathology.

Cite

CITATION STYLE

APA

Baron, J. M., Mermel, C. H., Lewandrowski, K. B., & Dighe, A. S. (2012). Detection of preanalytic laboratory testing errors using a statistically guided protocol. American Journal of Clinical Pathology, 138(3), 406–413. https://doi.org/10.1309/AJCPQIRIB3CT1EJV

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free