Abstract
In endovascular aneurysm repair (EVAR) procedures, the stent graft navigation and implantation is currently performed under a two-dimensional (2D) imaging based guidance requiring X-rays and contrast agent. In this work, a novel 3D stent graft guidance approach based on tracking systems is introduced. A calibration method and the visualization of the stent graft guidance are described. The tracking based stent graft guidance is evaluated by conducting an EVAR procedure on a torso phantom using a stent graft system equipped with an optical fiber and three EM sensors. The physicians were able to navigate the stent graft to the landing zone, and to place and implant it as intended using the introduced guidance. This showed that the application of the stent graft guidance is feasible in a clinical environment and promising for the reduction of radiation and contrast agent.
Author supplied keywords
Cite
CITATION STYLE
Jäckle, S., Eixmann, T., Matysiak, F., Sieren, M. M., Horn, M., Schulz-Hildebrandt, H., … Pätz, T. (2021). 3D Stent Graft Guidance based on Tracking Systems for Endovascular Aneurysm Repair. In Current Directions in Biomedical Engineering (Vol. 7). Walter de Gruyter GmbH. https://doi.org/10.1515/cdbme-2021-1004