A sealed water calorimeter for measuring absorbed dose

  • Domen S
N/ACitations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

The NIST sealed water calorimeter is intended for direct measurement of absorbed dose to water. This calorimeter was used for a series of approximately 3700 measurements to investigate the so-called heat defect, that is, anomalous endothermic or exothermic effects caused by dissolved gases. The three systems investigated were ''high-purity'' water saturated with N2, H-2, and mixtures of H-2/O2. The repeatability of the measurements of and dose rates for the Co-60 teletherapy beam was studied with different water fillings and accumulated absorbed dose. Measurements with the H-2/O2 system varied with accumulated absorbed dose. Based on the measurements and theoretical considerations, it appears that the H-2-saturated system is the best choice for eliminating the heat defect. Measurements with both the N2- and H-2-saturated systems are in good agreement with those determined with a graphite and graphite-water calorimeter (for which there is no heat defect).