Purpose: Short-stem implants provide a bone-preserving alternative in total hip arthroplasty. However, some evidence exists that the smaller implant-bone contact surface may compromise primary stability and impair osseo-integration. The purpose of this study was to analyse the migration characteristics of the Nanos® short stem over two years by means of model-based roentgen stereophotogrammetric analysis (MBRSA). Methods: Eighteen patients aged 53.6 ± 7.2 years were included. After being treated with a Nanos implant, 14 patients were followed-up radiologically at three, six, 12 and 24 months by means of MBRSA. Early implant migration was calculated. Clinical data have been assessed in addition. Results: Highest translational migration was observed with a mean value of –0.22 ± 0.39 mm along the proximo-distal axis after three months and highest rotational migration with 0.8 ± 3.2° also around the y-axis after two years. The resulting total migration was 0.46 ± 0.31 mm, with the largest proportion occurring within three months after surgery (0.40 ± 0.34 mm). Conclusion: The Nanos short-stem hip implant shows only a slight initial migration within three months after implantation, followed by secondary stabilisation. These results suggest both good primary stability and osseo-integration, suggesting a low risk of aseptic loosening.
Author supplied keywords
Cite
CITATION STYLE
Budde, S., Seehaus, F., Schwarze, M., Hurschler, C., Floerkemeier, T., Windhagen, H., … Thorey, F. (2016). Analysis of migration of the Nanos® short-stem hip implant within two years after surgery. International Orthopaedics, 40(8), 1607–1614. https://doi.org/10.1007/s00264-015-2999-9
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.