Abstract

A high-speed steady-state visual evoked potentials (SSVEP)-based brain-computer interface (BCI) system using dry EEG electrodes was demonstrated in this study. The dry electrode was fabricated in our laboratory. It was designed as claw-like structure with a diameter of 14 mm, featuring 8 small fingers of 6 mm length and 2 mm diameter. The structure and elasticity can help the fingers pass through the hair and contact the scalp when the electrode is placed on head. The electrode was capable of recording spontaneous EEG and evoked brain activities such as SSVEP with high signal-to-noise ratio. This study implemented a twelve-class SSVEP-based BCI system with eight electrodes embedded in a headband. Subjects also completed a comfort level questionnaire with the dry electrodes. Using a preprocessing algorithm of filter bank analysis (FBA) and a classification algorithm based on task-related component analysis (TRCA), the average classification accuracy of eleven participants was 93.2% using 1-second-long SSVEPs, leading to an average information transfer rate (ITR) of 92.35 bits/min. All subjects did not report obvious discomfort with the dry electrodes. This result represented the highest communication speed in the dry-electrode based BCI systems. The proposed system could provide a comfortable user experience and a stable control method for developing practical BCIs.

Cite

CITATION STYLE

APA

Xing, X., Wang, Y., Pei, W., Guo, X., Liu, Z., Wang, F., … Chen, H. (2018). A High-Speed SSVEP-Based BCI Using Dry EEG Electrodes. Scientific Reports, 8(1). https://doi.org/10.1038/s41598-018-32283-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free