Background: Closed-loop (CL) insulin delivery systems utilizing proportional-integral-derivative (PID) controllers have demonstrated susceptibility to late postprandial hypoglycemia because of delays between insulin delivery and blood glucose (BG) response. An insulin feedback (IFB) modification to the PID algorithm has been introduced to mitigate this risk. We examined the effect of IFB on CL BG control. Methods: Using the Medtronic ePID CL system, four subjects were studied for 24 h on PID control and 24 h during a separate admission with the IFB modification (PID + IFB). Target glucose was 120 mg/dl; meals were served at 8:00 AM, 1:00 PM, and 6:00 PM and were identical for both admissions. No premeal manual boluses were given. Reference BG excursions, defined as incremental glucose rise from premeal to peak, and postprandial BG area under the curve (AUC; 0-5 h) were compared. Results are reported as mean ± standard deviation. Results: The PID + IFB control resulted in higher mean BG levels compared with PID alone (153 ± 54 versus 133 ± 56 mg/dl; p < 60 mg/dl) occurred during PID control versus none during PID + IFB. Conclusions: Addition of IFB to the PID controller markedly reduced the occurrence of hypoglycemia without increasing meal-related glucose excursions. Higher average BG levels may be attributable to diferences in the determination of system gain (Kp) in this study. The prevention of postprandial hypoglycemia suggests that the PID + IFB algorithm may allow for lower target glucose selection and improved overall glycemic control. © Diabetes Technology Society.
CITATION STYLE
Ruiz, J. L., Sherr, J. L., Cengiz, E., Carria, L., Roy, A., Voskanyan, G., … Weinzimer, S. A. (2012). Effect of insulin feedback on closed-loop glucose control: A crossover study. Journal of Diabetes Science and Technology, 6(5), 1123–1130. https://doi.org/10.1177/193229681200600517
Mendeley helps you to discover research relevant for your work.