Abstract
Introduction: Orthopedic drilling and screw placement require precision to obtain the effect of bicortical screwing and avoid complications such as soft tissue damage. Traditional manual depth gauges, while effective, are prone to human error. This study evaluates the usability and acceptability of an automated electronic depth gauge compared to a conventional manual depth gauge in a preclinical setting. Methods: A randomized controlled cross-over trial was conducted with 45 participants of varying clinical experience (12 medical students, 11 resident doctors, and 22 consultants, 14/22 were senior consultants) at Aarhus University Hospital. According to randomization, participants used either the “electronic -> manual” or “manual -> electronic” depth gauge and subsequently the sequence “with -> without” or “without -> with” soft tissue protector to measure screw length in sawbones. Efficiency was assessed by measuring time (savings), and usability was evaluated using the System Usability Scale (SUS). Results: The electronic depth gauge significantly reduced the time required to measure screw length, with medical students saving the most time, p < 0.001. 65/331 (20%) of screws were changed, with no difference in the rate of screw changes between manual and electronic depth gauges, p = 0.76. Level of experience was significantly correlated with the rate of screw changes, p = 0.03. Resident doctors changed fewer screws than medical students, with no difference between the remaining groups. The median SUS score was 87.5, indicating excellent usability. Most participants rated the tool as “best imaginable” or “excellent.” Conclusion: The electronic depth gauge was time-efficient and surgeons rated it as highly usable. However, screw exchanges were not significantly minimized in this study, which might be mitigated by proper training and awareness of device-specific recommendations. The effect of electronic depth gauges on screw exchange rates and its clinical applicability warrents further research.
Author supplied keywords
Cite
CITATION STYLE
Engbjerg, J. S., Thorninger, R., Nielsen, C. L., Rinne, J., & Rölfing, J. D. (2025). Usability and acceptability of a power tool with electronic depth gauge for orthopedic drilling – a preclinical randomized controlled trial in sawbones. Archives of Orthopaedic and Trauma Surgery, 145(1). https://doi.org/10.1007/s00402-025-05839-3
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.