Direct screening of urine for MDMA and MDA by liquid chromatography-tandem mass spectrometry

33Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

There is a limit in the number of substances detected by commercially available reagents. It is therefore important to have other, complementary techniques. Liquid chromatography-tandem mass spectrometry (LC-MS-MS) may offer one possibility. An LC-MS-MS procedure based on the detection of positive ions after atmospheric pressure chemical ionization (APCI) was developed for direct measurement of 3,4-methylenedioxymethamphetamine (MDMA) and 3,4-methylenedioxyamphetamine in urine. It was compared with Online® Amphetamines. The LC-MS-MS methodology showed improved sensitivity (100-ng/mL cutoff) and specificity with a coefficient of variation of 10%. It was compared to the immunochemical analysis using 1000 clinical patient urine samples. The positive samples were confirmed by gas chromatography-mass spectrometry. The LC-MS-MS method detected almost four times as many samples positive for MDMA compared to the immunochemical method, with no false positives and one false negative. Our study suggests that LC-MS-MS offers an alternative to immunochemical methods in drug of abuse screening.

Cite

CITATION STYLE

APA

Nordgren, H. K., & Beck, O. (2003). Direct screening of urine for MDMA and MDA by liquid chromatography-tandem mass spectrometry. Journal of Analytical Toxicology, 27(1), 15–19. https://doi.org/10.1093/jat/27.1.15