Evaluation of a transportable capnometer for monitoring end-tidal carbon dioxide

16Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We compared a small and transportable Capnometer (EMMA™) with a reference capnometer, the Siesta i TS Anaesthesia. During air-breathing through a facemask, both the EMMA (nine modules) and reference capnometer sampled expired gas simultaneously. A wide range of end-tidal carbon dioxide values were obtained during inhalation of carbon dioxide and voluntary hyperventilation. The median IQR [range] difference between all sets of carbon dioxide values (EMMA - reference) was -0.3 (-0.6 to 0.0 [-1.7 to 1.6] kPa; n = 297) using new batteries, which was statistically significant (p = 0.04) and located to two of the nine EMMAs tested. Using batteries with reduced voltage did not influence the measurements. The 95% CI of the medians of the differences were -0.4 to -0.2. We conclude that the EMMA can slightly under-read the end-tidal carbon dioxide but is generally comparable with a free-standing monitor. The precision of the EMMAs was similar whether new batteries or batteries with reduced voltage were used. © 2010 The Association of Anaesthetists of Great Britain and Ireland.

Cite

CITATION STYLE

APA

Hildebrandt, T., Espelund, M., & Olsen, K. T. (2010). Evaluation of a transportable capnometer for monitoring end-tidal carbon dioxide. Anaesthesia, 65(10), 1017–1021. https://doi.org/10.1111/j.1365-2044.2010.06499.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free