Abstract
We compared a small and transportable Capnometer (EMMA™) with a reference capnometer, the Siesta i TS Anaesthesia. During air-breathing through a facemask, both the EMMA (nine modules) and reference capnometer sampled expired gas simultaneously. A wide range of end-tidal carbon dioxide values were obtained during inhalation of carbon dioxide and voluntary hyperventilation. The median IQR [range] difference between all sets of carbon dioxide values (EMMA - reference) was -0.3 (-0.6 to 0.0 [-1.7 to 1.6] kPa; n = 297) using new batteries, which was statistically significant (p = 0.04) and located to two of the nine EMMAs tested. Using batteries with reduced voltage did not influence the measurements. The 95% CI of the medians of the differences were -0.4 to -0.2. We conclude that the EMMA can slightly under-read the end-tidal carbon dioxide but is generally comparable with a free-standing monitor. The precision of the EMMAs was similar whether new batteries or batteries with reduced voltage were used. © 2010 The Association of Anaesthetists of Great Britain and Ireland.
Cite
CITATION STYLE
Hildebrandt, T., Espelund, M., & Olsen, K. T. (2010). Evaluation of a transportable capnometer for monitoring end-tidal carbon dioxide. Anaesthesia, 65(10), 1017–1021. https://doi.org/10.1111/j.1365-2044.2010.06499.x
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.