The impact of changing ventilator parameters on availability of nebulized drugs in an in vitro neonatal lung system

9Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

An in vitro model was developed to assess nebulized drug delivery. The model simulated the intubated neonate and examined the effect of changes in a variety of parameters commonly confronted in the clinical setting. Theophylline was nebulized for 15 minutes and captured in an artificial lung system (a 1000-mL intravenous bag). Variables were: Peak pressure (20, 24, 28 cm H2O), ventilator rate (40, 60, 80 breaths/min), nebulizer flow rate (5, 7, 10 L/min), endotracheal tube size (2.5, 3.0, 3.5 mm), and ventilator type (Servo 900C, Bourns BP 200, Bear Cub BP 2001). The amount of drug actually captured in the bag ranged from 0.009 to 12.59 percent (mean 2.08). A multivariate analysis showed that only nebulizer flow rate had a statistically significant effect on drug delivery with 10 L/min delivering the most drug. All factors combined only accounted for 11.5 percent of the variability in drug delivery. In light of the wide and unpredictable amounts of drug delivered through ventilators, dosing to pharmacologic effect rather than staying within narrow dosing guidelines may be more rational in patients responding poorly to standard doses. © 1991, SAGE Publications. All rights reserved.

Cite

CITATION STYLE

APA

Benson, J. M., Benson, J. M., Gal, P., Kandrotas, R. J., Watling, S. M., & Hansen, C. J. (1991). The impact of changing ventilator parameters on availability of nebulized drugs in an in vitro neonatal lung system. Annals of Pharmacotherapy, 25(3), 272–275. https://doi.org/10.1177/106002809102500311

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free