Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode

15Citations
Citations of this article
44Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: Through the Monte Carlo (MC) simulation of 6 and 10 MV flattening-filter-free (FFF) beams from Varian TrueBeam accelerator, this study aims to find the best incident electron distribution for further studying the small field characteristics of these beams. Methods: By incorporating the training materials of Varian on the geometry and material parameters of TrueBeam Linac head, the 6 and 10 MV FFF beams were modelled using the BEAMnrc and DOSXYZnrc codes, where the percentage depth doses (PDDs) and the off-axis ratios (OARs) curves of fields ranging from 4 × 4 to 40 × 40cm2 were simulated for both energies by adjusting the incident beam energy, radial intensity distribution and angular spread, respectively. The beam quality and relative output factor (ROF) were calculated. The simulations and measurements were compared using Gamma analysis method provided by Verisoft program (PTW, Freiburg, Germany), based on which the optimal MC model input parameters were selected and were further used to investigate the beam characteristics of small fields. Results: The Full Width Half Maximum (FWHM), mono-energetic energy and angular spread of the resultant incident Gaussian radial intensity electron distribution were 0.75mm, 6.1MeV and 0.9° for the nominal 6 MV FFF beam, and 0.7mm, 10.8MeV and 0.3° for the nominal 10 MV FFF beam respectively. The simulation was mostly comparable to the measurement. Gamma criteria of 1mm/1% (local dose) can be met by all PDDs of fields larger than 1 × 1cm2, and by all OARs of no larger than 20 × 20cm2, otherwise criteria of 1mm/2% can be fulfilled. Our MC simulated ROFs agreed well with the measured ROFs of various field sizes (the discrepancies were less than 1%), except for the 1 × 1cm2 field. Conclusions: The MC simulation agrees well with the measurement and the proposed model parameters can be clinically used for further dosimetric studies of 6 and 10 MV FFF beams.

Cite

CITATION STYLE

APA

Feng, Z., Yue, H., Zhang, Y., Wu, H., Cheng, J., & Su, X. (2016). Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode. Radiation Oncology, 11(1). https://doi.org/10.1186/s13014-016-0601-2

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free