Analytical performance of an inductively coupled plasma optical emission spectrometry with dual view configuration

23Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.

Abstract

The analytical performance of axially and radially viewed inductively coupled plasma optical emission spectrometers (AX-ICP OES and RD-ICP OES, respectively) were evaluated in terms of the Mg II/Mg I ratio, matrix effects, and detection limits (LOD). All solutions were made in either 1% v v-1 nitric acid or in a 10% v v-1 water-soluble tertiary amine solution (CFA-C) adjusted to pH 8. Using a nebulization gas flow-rate of 0.5 L min-1 and applied power of 1.3 kW, Mg II/Mg I ratio was at least 15 in any medium for both configurations. Under robust conditions, plasma processes were practically not affected by the presence of 0.1% m v-1 of Ca, K, or Na in both configurations. However, interference effects were observed with non-robust conditions, mainly with AX-ICP OES in both media. The best detection limits were obtained using AX-ICP OES under robust conditions, with solutions prepared in the amine medium.

Cite

CITATION STYLE

APA

Silva, J. C. J., Baccan, N., & Nóbrega, J. A. (2003). Analytical performance of an inductively coupled plasma optical emission spectrometry with dual view configuration. Journal of the Brazilian Chemical Society, 14(2), 310–315. https://doi.org/10.1590/S0103-50532003000200020