Abstract
The subcutaneous implantable cardioverter–defibrillator (S‐ ICD ) is a safe alternative to transvenous ICD . We describe a submuscular S‐ ICD placement technique in a severely obese with an oversized chest. Submuscular configuration allows optimal system positioning and impendence values warranting a safe and effective shock transmission. This technique is safe and improves patients comfort.
Cite
CITATION STYLE
Droghetti, A., Locatelli, A., Casiraghi, B., Malacrida, M., Arupi, M., & Ragusa, M. (2016). Totally submuscular implantation of subcutaneous implantable cardioverter defibrillator: a safe and effective solution for obese or oversized patients. Clinical Case Reports, 4(10), 1009–1011. https://doi.org/10.1002/ccr3.652