Abstract
Introduction Spinal anesthesia involves the administration of a dose of local anesthesia into the intrathecal space of the lumbar spine, using spinal needles with varying characteristics and design. When completing the administration of spinal anesthesia, you may check the volume of fluid remaining inside the spinal needle. Objective To measure and compare the volume of residual fluid following the administration of spinal anesthesia using 3 different brands of 27 G spinal needles. Methods This is an observational analytical cross-section trial for which 3 types of 27 G commercial needles were selected: BD Whitacre pencil point, Spinocan Quincke sharp point, and Disposable spinal needle. Following the administration of spinal anesthesia, the same operator, without detaching the syringe from the needle, deposited the residual volume into a collection tube so that a second operator quantifies the volume using a precision micropipette. Results The residual volume obtained from the 27 G spinal needles showed a statistically significant difference in favor of the BD Whitacre needles (P
Author supplied keywords
Cite
CITATION STYLE
Luna, C. A., Tulcán-Toro, R., Romero, F., & Luna, M. F. (2017). Measurement of residual volume in spinal needles after spinal anesthesia. Revista Colombiana de Anestesiologia, 45, 12–15. https://doi.org/10.1097/01819236-201712002-00003