Abstract
In this article, we investigate quantum entanglement (QE) filtering to address the challenges in multi-isotope positron emission tomography (PET) or in PET studies utilizing radiotracers with dual-positron and prompt gamma emissions. Via GATE simulation, we demonstrate the efficacy of QE filtering using a one-of-a-kind cadmium-zinc-telluride (CZT) PET system-establishing its viability as a multimodal scanner and ability to perform QE filtering. We show the high Compton scattering probability in this CZT-based scanner with 44.2% of gammas undergoing a single scatter prior to absorption. Additionally, the overall system sensitivity as a standard PET scanner (11.29%), QE-PET scanner (6.81%), and Compton camera (10.05%) is quantified. Further, we find a 23% decrease in the double Compton scatter (DCSc) frequency needed for QE filtering for each mm decrease in crystal resolution and an increase in mean absolute error (MAE) of their Δφs from 6.8° for 1 mm resolution to 9.5°, 12.2°, and 15.3° for 2, 4, and 8 mm resolution, respectively. These results reinforce the potential of CZT detectors to lead next-generation PET systems by fully leveraging QE information of positron annihilation photons.
Author supplied keywords
Cite
CITATION STYLE
Romanchek, G., Shoop, G., Gholami, K., Enlow, E., & Abbaszadeh, S. (2024). Quantum Entanglement Filtering: A PET Feasibility Study in Imaging Dual-Positron and Prompt Gamma Emission via Monte Carlo Simulation. IEEE Transactions on Radiation and Plasma Medical Sciences, 8(8), 916–925. https://doi.org/10.1109/TRPMS.2024.3388872