A modified simplex algorithm was used to optimize a system of preconcentration for cadmium determination in samples of sediments, cigarettes and food using flame atomic absorption spectrometry. The preconcentration system is based on the sorption of cadmium in a minicolumn packed with Amberlite XAD-2 resin functionalized with 3,4-dihydroxybenzoic acid (DHB). The optimized variables were pH and sampling flow rate and the optimum conditions found for these variables were, respectively 8.7 and 8.8 mL min-1. The developed system showed a preconcentration factor of 15.3, detection limit of 0.49 μg L-1, quantification limit of 1.65 μg L-1 and precision expressed as relative standard deviation (% RSD, n=10) of 5.9. The accuracy of the method was checked by analysis of estuary sediment certified reference material (NIST 1646-1). The cadmium concentrations found in sediment samples ranged from 1.73 and 1.90 μg g-1. In cigarette samples the results were 0.085 and 0.193 μg g-1, and in food samples (coriander and lettuce) the concentrations found of this metal were, respectively, 0.33 and 0.12 μg g-1.
Author supplied keywords
Cite
CITATION STYLE
dos Santos, Q. O., Moreno, I., dos Santos, L., Santos, A. G., Souza, V. S., & Bezerra, M. A. (2016). Application of modified simplex on the development of a preconcentration system for cadmium determination in sediments, food and cigarettes. Anais Da Academia Brasileira de Ciencias, 88(2), 791–799. https://doi.org/10.1590/0001-3765201620150209