Abstract

An immunoassay performed on a portable microfluidic device was evaluated for the determination of urinary albumin. An increase in absorbance at 500 nm resulting from immunoagglutination was monitored directly on the poly(dimethylsiloxane) (PDMS) microchip using a portable miniature fibre-optic spectrometer. A calibration curve was linear up to 10 mg L-1 (r2 = 0.993), with a detection limit of 0.81 mg L-1 (S/N = 3). The proposed system showed good precision, with relative standard deviations (RSDs) of 5.1%, when evaluated with 10 mg L-1 albumin (n = 10). Determination of urinary albumin with the proposed system gave results highly similar to those determined by the conventional spectrophotometric method using immunoturbidimetric detection (r2 = 0.995; n = 15). © 2009 by the authors.

Cite

CITATION STYLE

APA

Laiwattanapaisal, W., Songjaroen, T., Maturos, T., Lomas, T., Sappat, A., & Tuantranont, A. (2009). On-chip immunoassay for determination of urinary albumin. Sensors, 9(12), 10066–10079. https://doi.org/10.3390/s91210066