Abstract
Purpose:To evaluate the long-term effectiveness and safety of the XtraFocus intraocular pinhole (IOPH) when it is implanted inside the capsular bag.Setting:Private practice.Design:Retrospective consecutive case series.Methods:Patients who had an IOPH implanted in the capsular bag together with the primary intraocular lens (IOL) to treat irregular corneal astigmatism secondary to multiple causes were enrolled. The mean follow-up was 16 months (range 7 to 48 months). Patients were assessed in their scheduled follow-up visits. The uncorrected and corrected distance visual acuities were recorded at each visit. An infrared slitlamp photograph was captured and analyzed to verify the presence of interlenticular membrane formation.Results:Sixty eyes of 58 patients were analyzed. The mean uncorrected and corrected distance visual acuities improved from logarithm of the minimum angle of resolution 1.34 ± 0.338 and 0.57 ± 0.145 preoperatively to 0.14 ± 0.012 (P
Cite
CITATION STYLE
Trindade, B. L. C., Trindade, F. C., Werner, L., & Trindade, C. L. C. (2020). Long-term safety of in-the-bag implantation of a supplementary intraocular pinhole. Journal of Cataract and Refractive Surgery, 46(6), 888–892. https://doi.org/10.1097/j.jcrs.0000000000000163