Abstract

Background Dynamic contrast-enhanced MRI (DCE-MRI) biomarkers have proven utility in tumors in evaluating microvascular perfusion and permeability, but it is unclear whether measurements made in different centers are comparable due to methodological differences. Purpose To evaluate how commonly utilized analytical methods for DCE-MRI biomarkers affect both the absolute parameter values and repeatability. Materials and Methods DCE-MRI was performed on three consecutive days in twelve rats bearing C6 xenografts. Endothelial transfer constant (K trans), extracellular extravascular space volume fraction (v e), and contrast agent reflux rate constant (k ep) measures were computed using: 2-parameter ("Tofts" or "standard Kety") vs. 3-parameter ("General Kinetic" or "extended Kety") compartmental models (including blood plasma volume fraction (v p) with 3-parameter models); individual- vs. population-based vascular input functions (VIFs); and pixel-by-pixel vs. whole tumor-ROI. Variability was evaluated by within-subject coefficient of variation (wCV) and variance components analyses. Results DCE-MRI absolute parameter values and wCVs varied widely by analyticalmethod. Absolute parameter values ranged, as follows, median K trans, 0.09-0.18 min -1; k ep, 0.51-0.92 min -1; v e, 0.17-0.23; and v p, 0.02-0.04. wCVs also varied widely by analytical method, as follows: mean K trans, 32.9-61.9%; k ep, 11.6-41.9%; v e, 16.1-54.9%; and v p, 53.9-77.2%. K trans and kep values were lower with 3- than 2-parameter modeling (p<0.0001); k ep and v p were lower with pixel- than whole-ROI analyses (p<0.0006). wCVs were significantly smaller for v e, and larger for k ep, with individual- than population-based VIFs. Conclusions DCE-MRI parameter values and repeatability can vary widely by analytical methodology. Absolute values of DCE-MRI biomarkers are unlikely to be comparable between different studies unless analyses are carefully standardized.

Cite

CITATION STYLE

APA

Ng, C. S., Wei, W., Bankson, J. A., Ravoori, M. K., Han, L., Brammer, D. W., … Jackson, E. F. (2015). Dependence of DCE-MRI biomarker values on analysis algorithm. PLoS ONE, 10(7). https://doi.org/10.1371/journal.pone.0130168

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free