Design of LabVIEW®‐based software for the control of sequential injection analysis instrumentation for the determination of morphine

  • Lenehan C
  • Barnett N
  • Lewis S
N/ACitations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

LabVIEW ® ‐based software for the automation of a sequential injection analysis instrument for the determination of morphine is presented. Detection was based on its chemiluminescence reaction with acidic potassium permanganate in the presence of sodium polyphosphate. The calibration function approximated linearity (range 5 × 10 −10 to 5 × 10 −6 M) with a line of best fit of y = 1.05 x + 89164( R 2 = 0.9959), where y is the log10 signal (mV) and x is the log10 morphine concentration (M). Precision, as measured by relative standard deviation, was 0.7% for five replicate analyses of morphine standard (5 × 10 −8 M). The limit of detection (3σ) was determined as 5 × 10 −11 M morphine.

Cite

CITATION STYLE

APA

Lenehan, C. E., Barnett, N. W., & Lewis, S. W. (2002). Design of LabVIEW®‐based software for the control of sequential injection analysis instrumentation for the determination of morphine. Journal of Analytical Methods in Chemistry, 24(4), 99–103. https://doi.org/10.1155/s1463924602000135

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free