Inoculum size and lot-to-lot variation as significant variables in the tube coagulase test for Staphylococcus aureus

8Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

For a positive coagulase test with certain isolates of Staphylococcus aureus, inoculation of 0.5 ml of coagulase plasma with at least two colonies and 24 h of incubation at 35 to 37°C may be required. S. aureus does not multiply in coagulase plasma. There is also lot-to-lot variation in the sensitivity of coagulase plasma. On isolates from critical specimens, a negative coagulase test obtained with an inoculum of less than two colonies should be confirmed by repeating the procedure with a larger inoculum or by using some other confirmatory test.