Abstract
This paper demonstrates multiphoton excited fluorescence imaging through a polarisation maintaining multicore fiber (PM-MCF) while the fiber is dynamically deformed using all-proximal detection. Single-shot proximal measurement of the relative optical path lengths of all the cores of the PM-MCF in double pass is achieved using a Mach-Zehnder interferometer read out by a scientific CMOS camera operating at 416 Hz. A non-linear least squares fitting procedure is then employed to determine the deformation-induced lateral shift of the excitation spot at the distal tip of the PM-MCF. An experimental validation of this approach is presented that compares the proximally measured deformation-induced lateral shift in focal spot position to an independent distally measured ground truth. The proximal measurement of deformation-induced shift in focal spot position is applied to correct for deformation-induced shifts in focal spot position during raster-scanning multiphoton excited fluorescence imaging.
Cite
CITATION STYLE
Warren, S. C., Kim, Y., Stone, J. M., Mitchell, C., Knight, J. C., Neil, M. A. A., … Dunsby, C. (2016). Adaptive multiphoton endomicroscopy through a dynamically deformed multicore optical fiber using proximal detection. Optics Express, 24(19), 21474. https://doi.org/10.1364/oe.24.021474