We made a clinical study of polyethylene wear in 240 hips of 187 patients having primary total hip arthroplasties from 1989 to 1990, using uncemented Osteonics components, with a head size of 26 mm. We excluded cups with anteversion of over 20°and measured linear wear by a new method using a digitiser and special software of our design. Follow-up was from two to five years (mean 4.3). The mean age at operation was 50.3 years, with more men than women (1.4:1). The mean linear wear per year was 0.15mm; this did not increase with the longevity of the prosthesis (p = 0.54). In 59 hips showing evidence of osteolysis, the mean linear wear rate was significantly higher at 0.23 mm/year (p < 0.001). The mean linear wear rate also correlated significantly with age at the time of operation (p = 0.008), but we found no significant correlations with body-weight, gender, aetiology of the disease, thickness of polyethylene, or cup position. Our new method of measurement is time-saving and reproducible. The results confirm the greater rate of linear wear of polyethylene in patients showing osteolysis and in those who are younger.
CITATION STYLE
Shih, C. H., Lee, P. C., Chen, J. H., Tai, C. L., Chen, L. F., Wu, J. S. S., & Chang, W. H. (1997). Measurement of polyethylene wear in cementless total hip arthroplasty. Journal of Bone and Joint Surgery - Series B, 79(3), 361–365. https://doi.org/10.1302/0301-620X.79B3.6630
Mendeley helps you to discover research relevant for your work.