Determination of niacin in food materials by liquid chromatography using isotope dilution mass spectrometry

15Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The availability of deuterium-labeled nicotinic acid makes stable isotope dilution mass spectromerty (MS) coupled with liquid chromatography (LC) an attractive option for the determination of the water-soluble B-vitamin niacin in food samples. A method was developed based on acid digestion, solid-phase extraction with a strong cation exchange column, and reversed-phase chromatography with a C18 column. Detection is by positive ion electrospray MS. Analysis in selected ion recording mode is subject to interference problems similar to those found with other LC determinations of niacin, but the additional selectivity of multiple reaction monitoring mode largely eliminates interference problems. The method was applied to 6 different food matrixes and to appropriate reference materials, including milk samples with niacin levels near 1 ppm. The method exhibited good accuracy, based on levels obtained for the reference materials, and relative standard deviations in the range of 0.5-5%.

Cite

CITATION STYLE

APA

Goldschmidt, R. J., & Wolf, W. R. (2007). Determination of niacin in food materials by liquid chromatography using isotope dilution mass spectrometry. Journal of AOAC International, 90(4), 1084–1089. https://doi.org/10.1093/jaoac/90.4.1084