A fast analytical method for determination of hydroquinone in pharmaceutical formulations employing batch injection analysis (BIA) with amperometric detection using a boron-doped diamond electrode is described. The supporting electrolyte was a 0.1 mol L-1 H2SO4 solution (the single reagent used for analysis). The method showed good repeatability (RSD of 0.45%, n=20), wide linear range (from 10 to 2000 μmol L-1, R=0.9999), low detection limit (0.016 μmol L-1) and satisfactory recovery values (91-96%). Accuracy of the method was evaluated by comparative analyses using high-performance liquid-chromatography. The ability to replace the electronic pipette by disposable syringes (injection procedure) in BIA systems was also shown.
CITATION STYLE
Cunha, R. R., Tormin, T. F., Richter, E. M., & Munoz, R. A. A. (2013). Determinação rápida de hidroquinona usando aná:lise por injeção em batelada (bia) com detecção amperométrica. Quimica Nova, 36(5), 663–668. https://doi.org/10.1590/S0100-40422013000500009