Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator

16Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

Abstract

Aims Inappropriate shocks (IAS) remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD). The aims were to assess and characterize different patterns of IAS. Methods Two hundred and thirty-nine patients were implanted with an S-ICD between 2010 and 2018 for primary and sec- and results ondary prevention. Follow-up data of at least 6 months were analysed. During a mean follow-up of 34.9 ± 16.0 months, a total of 73 shocks occurred in 38 patients (6%). Forty-three (59%) shocks were considered appropriate due to ventricular tachycardia/ventricular fibrillation, while 30 (41%) were inappropriate and occurred in 19 patients (8%). Myopotentials/noise was the most frequent cause of inappropriate shocks (n = 8), followed by T-wave oversensing (n = 6) and undersensing of the QRS, resulting in adaptation of the automatic gain control and inappropriate shock (n = 5). Seventy-four percent of all IAS occurred on the primary vector, while no IAS occurred on the alternate vector. In seven of eight patients (88%), IAS related to myopotentials have occurred on the primary sensing vector. Multivariate analysis identified taller patients, primary sensing vector and first-generation S-ICD device as predictors for IAS. SMART pass effectively reduced the occurrence of IAS in the second-generation S-ICD system. Conclusion Inappropriate therapies are less frequently observed on the alternate vector. The primary vector seems to be unfavourable with regard to oversensing caused by myopotentials. Inappropriate shocks were associated with an increased rate of rehospitalization but not mortality. These observations have implications for the prevention of inappropriate S-ICD shocks.

Cite

CITATION STYLE

APA

Rudic, B., Tülümen, E., Fastenrath, F., Röger, S., Goranova, D., Akin, I., … Kuschyk, J. (2020). Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator. Europace, 22(5), 761–768. https://doi.org/10.1093/europace/euaa026

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free