Abstract
High performance liquid chromatography (HPLC) is increasingly being used to estimate variant haemoglobins. A case of haemoglobin S/D (HbS/D) is presented, which was misdiagnosed as haemoglobin S/E (HbS/E) by HPLC. The patient was a 22 year old woman with sickle cell anaemia. Subsequent haemodilution by blood transfusion clearly elucidated the haemoglobin D peak on HPLC. Sample dilution experiments, using the pretransfusion sample, were done resulting in correct elution of the peak in the D window. Troubleshooting in similar problematic haemoglobin variant peaks seen on HPLC can be done by sample dilution.
Author supplied keywords
Cite
CITATION STYLE
Thomas, S. (1998). Sample dilution to resolve mistaken identification of haemoglobin D as haemoglobin E using the variant automated system. Journal of Clinical Pathology, 51(3), 253–254. https://doi.org/10.1136/jcp.51.3.253