Clinical diagnosis of fissure caries with conventional and laser-induced fluorescence techniques

46Citations
Citations of this article
62Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We studied the in vivo validity of dentinal fissure caries diagnosis by visual examination, bitewing radiography, and use of a laser-induced fluorescence device (DIAGNOdent). A total of 144 and second molars with macroscopically intact occlusal surfaces in 41 Chinese young adults were examined visually, by bitewing radiography, and by DIAGNOdent. Visual examination after pit and fissure opening was used as the reference standard. The sensitivity and specificity of detecting caries that had extended into the dentin were, respectively, 0.89 and 0.44 by visual detection of opacity or discoloration after air drying, 0.13 and 1.00 by bitewing radiography to detect radiolucency extending into the dentin, and 0.70 and 0.84 by DIAGNOdent testing with a cut-off score of 40. Caries detection by a combination of visual examination and DIAGNOdent had a sensitivity of 0.67 and specificity of 0.94. Receiver operating characteristic analysis showed that this combined approach was superior to the other methods. © Springer-Verlag London Limited 2009.

Cite

CITATION STYLE

APA

Chu, C. H., Lo, E. C. M., & You, D. S. H. (2010). Clinical diagnosis of fissure caries with conventional and laser-induced fluorescence techniques. Lasers in Medical Science, 25(3), 355–362. https://doi.org/10.1007/s10103-009-0655-6

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free