Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery

103Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The potential problems of monopolar electrosurgery relate to unrecognized energy transfer ("stray current") outside the view of the laparoscope. Mechanisms of stray current and uncrecognized tissue injury include: (1) insulation breaks in electrodes; (2) capacitive coupling, or induced currents through the intact insulation of the active electrode to surrounding cannulas or other instruments; and (3) direct coupling (or unintended contact) between the active electrode and other metal instruments or cannulas within the abdomen. Capacitive coupling poses the greatest risk for injury when the outer conductor (trocar cannula or irrigation cannula) is electrically isolated from the abdominal wall by a plastic nonconductor. Capacitive coupling is increased by the coagulation mode (versus cut), open circuit (versus tissue contact with the electrode), 5-mm cannulas (versus 11 mm), and higher voltage generators. The safety of electrosurgery can be enhanced by surgical education regarding the biophysics of radio frequency electrical energy, technical choices in instruments using all-metal cannula systems, and engineering developments with a dynamically monitored system for insulation failure and capacitive coupling. © 1992 Reed Publishing USA.

Cite

CITATION STYLE

APA

Voyles, C. R., & Tucker, R. D. (1992). Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. The American Journal of Surgery, 164(1), 57–62. https://doi.org/10.1016/S0002-9610(05)80648-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free