Evaluation of the Optilite® analyser for determination of total complement activity and C3 and C4 fractions

2Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

The complement system is composed of a set of plasma or membrane proteins. Complement protein deficiencies can be inherited or acquired, through the presence of autoantibodies or through consumption. We evaluated the analytical performance of the Optilite® analyser for the determination of the C3 and C4 levels and for the evaluation of the total complement activity. The intra- and inter-series CVs were evaluated and have showed satisfactory results, the concordances with analysers currently used in the laboratory (BNII® and BCT®, Siemens) are very good, as is the agreement between the serum and plasma samples. We also determined the reference values for the different parameters tested in viewof a routine use of Optilite® analyser in the laboratory.