Verification of automated latex-enhanced particle immunoturbidimetric D-dimer assays on different analytical platforms and comparability of test results

11Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

Introduction: The aim of the study was the analytical verification of automated latex-enhanced particle immunoturbidimetric (LPIA) D-Dimer assay INNOVANCE D-dimer on Sysmex CS-5100 and Atellica COAG 360 analysers, and HemosIL D-dimer HS500 on ACL TOP 550, as well as the comparison with the enzyme-linked immunofluorescent assay (ELFA) on the miniVidas analyser. Materials and methods: Verification included assessment of within-run and between-run precision, bias, measurement uncertainty (MU), verification of the cut-off, method comparison between all assessed assays, and the reference commercial ELFA VIDAS D-Dimer Exclusion II. Results: Within-run coefficients of variations (CVs) ranged from 1.6% (Atellica COAG 360) to 7.9% (ACL TOP 550), while between-run CVs ranged from 1.7% (Sysmex CS-5100) to 6.9% (Atellica COAG 360). Spearman’s rank correlation coefficients were > 0.99 between LPIAs and ≥ 0.93 when comparing ELFA with LPIA. Passing-Bablok regression analysis yielded constant and proportional difference for comparison of ACL TOP 550 with both Sysmex CS-5100 and Atellica COAG360, and for miniVidas with Atellica COAG360. Small proportional difference was found between miniVidas and both Sysmex CS-5100 and ACL TOP 550. Calculated MUs using D-dimer HS 500 calibrator were 12.6% (Sysmex CS-5100) and 15.6% (Atellica COAG 360), while with INNOVANCE D-dimer calibrator 12.0% (Sysmex CS-5100), 10.0% (Atellica COAG 360) and 28.1% (ACL TOP 550). Excellent agreement of results was obtained, with occasional discrepancies near the cut-off. The cut-off (0.5 mg/L FEU) was confirmed. Conclusions: The obtained results prove satisfactory analytical performance of LPIAs, their high comparability and almost equal discriminatory characteristics, suggesting them as a valid alternative to ELFA.

Cite

CITATION STYLE

APA

Lapić, I., Herak, D. C., Prpić, S., Prce, A., Raščanec, V., Zadro, R., & Rogić, D. (2020). Verification of automated latex-enhanced particle immunoturbidimetric D-dimer assays on different analytical platforms and comparability of test results. Biochemia Medica, 30(3), 1–9. https://doi.org/10.11613/BM.2020.030705

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free