Benefits and medium-term outcome of the Sorin Pericarbon Freedom stentless aortic prosthesis in cases of acute bacterial endocarditis

2Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

OBJECTIVES The aim of this study is to evaluate the ease of use and the advantages of Sorin Pericarbon Freedom (SPF) stentless valve in cases of acute bacterial endocarditis and to check the intermediate-term results after the implant of SPF with respect to resistance to infection, valve deterioration and durability. METHODS Between June 2003 and February 2015, 26 patients with active aortic valve bacterial endocarditis underwent aortic valve replacement with SPF pericardial stentless aortic prosthesis. The mean age was 57 ± 18 years; 73% of the patients were in preoperative NYHA class III and VI. Mean Logistic EuroSCORE was 14.2 ± 12.7. Endocarditis occurred in 18 patients with native valves, and in 9 patients with prosthetic valves (4 mechanical aortic valve prostheses; 5 aortic bioprostheses). Aortic root abscesses were observed in 16 cases (61.5%). Surgery was emergent in 3 cases (11.5%). Redo surgery was performed in 9 cases (35%). Cumulative follow-up was 126.8 patient-years (mean 4.9 ± 3.3 years). RESULTS Operative hospital mortality was 0% for all patients. Residual mean prosthetic gradient at discharge was 9.4 ± 3.6 mmHg. Neither residual aortic incompetence nor residual abscess cavity was observed at discharge. Mean ejection fraction at discharge was 54 ± 8% (Min; Max: 35%; 65%). A total of 4 patients died at follow-up, all for non-cardiac causes. One patient was lost to follow-up. Two patients (8%) underwent non-valve-related reoperation with 0% mortality. Residual mean gradient at follow-up was 7.2 ± 2.1 mmHg. Three patients (17%) presented with mild/moderate aortic incompetence and 89% of patients were in NYHA Class I-II at follow-up. At 9 years, actuarial freedom from valve-related reoperation and from structural valve deterioration was 100%. CONCLUSIONS The SPF aortic prosthesis is a true pericardial stentless prosthesis suitable for the treatment of acute bacterial endocarditis. Intermediate-time results in terms of freedom from reoperation, structural valve deterioration and resistance to infections are satisfactory. Haemodynamic performances are excellent since a complete exclusion of aortic root abscesses is achieved without any reduction of the aortic annular diameter, usually due to marsupialization or patch closure of the infected cavities.

Cite

CITATION STYLE

APA

Stefanelli, G., Pirro, F., Meli, M., Trevisan, D., Telesca, M., Campisano, B., … Barbieri, A. (2016). Benefits and medium-term outcome of the Sorin Pericarbon Freedom stentless aortic prosthesis in cases of acute bacterial endocarditis. Interactive Cardiovascular and Thoracic Surgery, 23(4), 593–598. https://doi.org/10.1093/icvts/ivw176

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free