Excessive pressure in multichambered cuffs used for sequential compression therapy

40Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background and Purpose. Pneumatic compression devices, used as part of the therapeutic strategy for lymphatic drainage, often have cuffs with multiple chambers that are inflated sequentially. The purpose of this study was to investigate (1) the relationship between cuff chamber pressure (Pchamber) and the pressure on the cuff-skin interface (Pinterface) and (2) the mechanical interaction of cuff chambers and consequences for device control. Subjects and Methods. In this study, we used 3 cylindrical (60-, 80-, and 100-mm-diameter) model limbs and 1 ellipsoidal model of the arm to test a commercially available pressure controller using "target pressures," indicated by the controller, of 30, 60, 80, and 100 mm Hg. We studied the time course of Pchamber and Pinterface during the inflation sequence and the effect of local curvature on Pinterface. Results. Our data indicated that, overall, Pinterface is of the same order of magnitude as Pchamber. There was some effect of model diameter and shape, with the smaller curvatures yielding the highest Pinterface. Cuff chamber interaction led to Pchamber and Pinterface values in the most distal (first inflated) chamber that were up to 80% higher than the target pressure. For the 80-mm cylindrical model, for instance, pressure in this chamber reached 54, 98, 121, and 141 mm Hg, respectively, instead of the 30, 60, 80, and 100 mm Hg indicated by the controller. Discussion and Conclusion. The discrepancy between the target pressure, indicated by the controller, and the pressure measured inside the cuff chambers undermines the therapeutic control and efficacy of the pneumatic compression devices. Because the measured pressures were far beyond the pressure level indicated by the controller, it is recommended that pneumatic compression devices be used at much lower target pressures (<30 mm Hg) than those applied in clinical practice.

Cite

CITATION STYLE

APA

Segers, P., Belgrado, J. P., Leduc, A., Leduc, O., & Verdonck, P. (2002). Excessive pressure in multichambered cuffs used for sequential compression therapy. Physical Therapy, 82(10), 1000–1008. https://doi.org/10.1093/ptj/82.10.1000

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free