Comparison of total ozone measurements in Melbourne, Australia, performed with a low-cost micro spectrometer and a Brewer MK-III

0Citations
Citations of this article
3Readers
Mendeley users who have this article in their library.

Abstract

A new design for a simple and very low-cost instrument for measuring total column ozone is described and its performance evaluated by comparison to a co-located MK III Brewer spectrophotometer for approximately 6 months. The ozone retrieval is based on the "Global Irradiance"method used to derive total column ozone from the Norwegian GUV radiometer network. While the total cost of components of the instrument was less than EUR 3000, total ozone values were found to agree with Direct Sun Brewer values with a standard deviation of 1.8 %, comparable to the agreement between DS and ZS Brewer data. The most significant limitations of the retrieval method were found to be the dependence on temperature and cloud-amount. As well as being low-cost, the instrument appears robust and easy to operate. As the instrument covers fully the UV-A/B spectrum it is fully possible to further develop the analyses software to include other data products related to UV-radiation.