Optimized goniometer for determination of the scattering phase function of suspended particles: simulations and measurements

  • Foschum F
  • Kienle A
34Citations
Citations of this article
44Readers
Mendeley users who have this article in their library.

Abstract

We present simulations and measurements with an optimized goniometer for determination of the scattering phase function of suspended particles. We applied the Monte Carlo method, using a radially layered cylindrical geometry and mismatched boundary conditions, in order to investigate the influence of reflections caused by the interfaces of the glass cuvette and the scatterer concentration on the accurate determination of the scattering phase function. Based on these simulations we built an apparatus which allows direct measurement of the phase function from ϑ=7  deg to ϑ=172  deg without any need for correction algorithms. Goniometric measurements on polystyrene and SiO2 spheres proved this concept. Using the validated goniometer, we measured the phase function of yeast cells, demonstrating the improvement of the new system compared to standard goniometers. Furthermore, the scattering phase function of different fat emulsions, like Intralipid, was determined precisely.

Cite

CITATION STYLE

APA

Foschum, F., & Kienle, A. (2013). Optimized goniometer for determination of the scattering phase function of suspended particles: simulations and measurements. Journal of Biomedical Optics, 18(8), 085002. https://doi.org/10.1117/1.jbo.18.8.085002