Abstract
Background: Safe and effective self-management of glucose levels requires immediate access to accurate data. We assessed the point accuracy of the Dexcom G7 Continuous Glucose Monitoring System (Dexcom, Inc., San Diego, CA, USA) and FreeStyle Libre 3 (Abbott Diabetes Care, Alameda, CA, USA) sensors in a head-to-head comparison. Method: Multicenter, single-arm, prospective, nonsignificant risk evaluation enrolled adults (≥ 18 years) with diagnosed type 1 diabetes (T1D) or type 2 diabetes (T2D). Accuracy was assessed by comparing sensor data to laboratory reference values Yellow Springs Instrument [YSI] and capillary blood glucose values. Outcome measures were differences in mean absolute relative difference (MARD), number and percentage of matched glucose pairs within ±20 mg/dL/±20 of reference values within glucose ranges: < 54, 54 to 69, 70 to 180, 181 to 250, > 250 mg/dL, and combined. Results: Data from 55 adults were included in the analysis. Analysis showed significantly lower MARD with the FreeStyle Libre 3 sensor vs the Dexcom G7 sensor (8.9% vs 13.6%, respectively, P
Author supplied keywords
Cite
CITATION STYLE
Hanson, K., Kipnes, M., & Tran, H. (2024). Comparison of Point Accuracy Between Two Widely Used Continuous Glucose Monitoring Systems. Journal of Diabetes Science and Technology, 18(3), 598–607. https://doi.org/10.1177/19322968231225676